Elemental Analysis Recovery: Expert Fixes for Low Results in CHNS, ICP-MS, ICP-OES, and AAS

This guide explains proven, lab-ready methods to diagnose and correct low recovery in elemental analysis so you can restore data accuracy with minimal reruns.

1. Define “Low Recovery” and Set Pass Criteria

Recovery is the measured amount of an analyte relative to its true or fortified amount, expressed as a percentage.

Spike Recovery (%) = 100 × (C_spiked − C_unspiked) / C_added. Standard Recovery (%) = 100 × C_measured / C_true.

Use method-specific acceptance windows. Typical goals are 90–110% for standards and 80–120% for matrix spikes unless the governing method states otherwise.

2. Isolate the Failure Mode With Fast QA/QC Checks

QC CheckPurposeInterpretation if Low RecoveryAction
Continuing calibration verification (CCV).Detect drift or bias.All matrices low.Recalibrate, verify fresh standards, check integration parameters.
Initial calibration linearity.Confirm calibration validity.Curvature or low slope.Reprepare standards, adjust ranges, use weighted regression.
Method blank.Detect contamination or carryover.Blank near zero but samples low.Bias is not from contamination. Inspect preparation or matrix effects.
Matrix spike (MS) and duplicate (MSD).Probe matrix effects.Low MS but good CCV.Matrix suppression or loss. Apply dilution, modifiers, or matrix match.
Post-digestion spike or post-combustion spike.Separate prep loss from instrumental suppression.Low post-spike.Instrument or solution chemistry issue.
Internal standards (metals) or control compounds (CHNS).Track response stability.Internal standards low.Nebulizer flow, plasma robustness, ionization effects, or leaks.

3. Root Causes and High-Yield Fixes by Technique

3.1 CHNS/O Combustion Analysis

SymptomLikely CauseCorrective Action
Consistently low C and H.Incomplete combustion or soot formation.Increase combustion temperature and oxygen pulse. Add vanadium pentoxide or tungsten trioxide catalyst. Reduce sample mass for high-carbon or high-ash matrices. Use tin capsules for better exotherm.
Low N with halogenated or nitro samples.Acid gases trapping reduced nitrogen species.Add combustion promoters. Use Ag or Cu oxide layers to capture halides. Optimize reduction furnace temperature for NOx to N2 conversion.
Low S in mineralized samples.S retained as sulfates in ash.Add WO3 accelerator. Increase oxygen flow. Ash sample separately and analyze filters if needed.
Memory between samples, then low recovery.Carryover or reagent trap saturation.Regenerate or replace absorber tubes. Run blanks between high-S or high-N samples. Verify water trap capacity.
Drift over sequence.Leaks or reagent degradation.Leak test. Replace combustion and reduction reagents. Recondition columns. Recalibrate with certified organics.
Caution: Halogen and sulfur-rich materials generate corrosive gases. Use appropriate traps, catalysts, and ventilation during combustion.

3.2 ICP-MS and ICP-OES for Metals

SymptomLikely CauseCorrective Action
Low recovery in all samples and CCV.Calibration drift, standard instability.Remake standards with high-purity acids and Type I water. Use single-use intermediate stocks. Verify unexpired CRM certificates.
Low recovery in matrices but CCV passes.Matrix suppression, high TDS, viscosity effects.Apply 2–10× dilution. Matrix-match standards. Add internal standards closely matched by mass and ionization energy. Use robust plasma (higher RF power, lower nebulizer gas). Use collision or reaction cell for polyatomic interferences.
Analyte-specific bias.Spectral overlap or ionization competition.Switch to alternate wavelengths or isotopes. Enable KED or reaction chemistry. Add ionization buffers like K or Na if allowed.
Post-digestion spike low but post-column spike OK.Deposits in spray chamber or nebulizer.Clean nebulizer and spray chamber. Verify peristaltic tubing compression and age. Confirm sample uptake rate with a stopwatch.
Instability and low internal standard signal.Air leak or salt buildup.Inspect pump tubing connections and drain. Check torch alignment, cones, and orifice cleanliness. Replace cones if worn.
Caution: HF digestion requires HF-compatible vessels and strict PPE. Do not introduce HF into glass components. Follow your acid neutralization SOP.

3.3 AAS (Flame and Graphite Furnace)

SymptomLikely CauseCorrective Action
Low flame AAS signal in viscous matrices.Poor nebulization and transport efficiency.Use matrix modifiers or dilute. Match acid and solvent composition of standards. Trim sample capillaries. Verify burner head cleanliness and alignment.
Low GFAAS recovery with refractory analytes.Formation of thermally stable oxides.Use chemical modifiers such as Pd or Mg(NO3)2. Optimize pyrolysis and atomization temperatures using standards in matrix match.
Baseline depression during read.Background absorption from matrix.Use Zeeman or deuterium background correction. Choose lines less prone to interference. Increase read delay after injection.

4. Sample Preparation: The Highest Leverage Step

4.1 Acid Digestion for Metals

Select a digestion chemistry that fully solubilizes the matrix while preserving analytes.

  • Silicates and refractory oxides. Use HF with HNO3, optionally HCl, in HF-compatible vessels. Complex residual fluoride with boric acid before analysis.
  • Organic matrices. Use HNO3 with H2O2. Add gentle heat or microwave digestion to complete oxidation.
  • High chloride matrices. Expect ArCl and related interferences in ICP-MS. Use collision cell or choose alternative isotopes.
  • Sulfide ores. Open vessels can lose volatile species. Use sealed microwave digestion to prevent loss.

4.2 Combustion for CHNS/O

  • Weigh 1–3 mg for high-C organics and 5–10 mg for inorganics or low-C matrices as permitted by instrument capacity.
  • Use tin capsules for energetic combustion. Add WO3 or V2O5 accelerators for difficult matrices.
  • Condition combustion and reduction reactors per manufacturer. Replace reagents on schedule.
  • Verify oxygen pulse duration and flow. Confirm carrier gas purity and leak-free lines.

5. Standards, Calibrants, and Controls

  • Use certified reference materials with current certificates. Record lot, preparation date, and storage conditions.
  • Prepare multi-element standards to bracket sample concentrations. Use weighted linear regression for wide ranges.
  • Matrix-match or use the method of standard additions when suppression persists.
  • For ICP-MS, assign internal standards near analyte masses. Typical mapping pairs include Sc, Ge, Rh, In, and Bi across the mass range.
  • For CHNS, run caffeine, acetanilide, or other certified organics as calibration and verification points.

6. Quick Decision Path to Fix Low Recovery

# Fast triage workflow 1. Check CCV. If low, recalibrate and verify standard integrity. 2. Run post-prep spike. If low, instrument or solution chemistry issue. 3. Run post-column spike (metals). If normal, sample introduction problem. 4. Compare MS/MSD. If low only in matrix, apply dilution or matrix match. 5. Inspect sample prep. Confirm complete digestion or combustion. 6. Adjust method. Alternate lines/isotopes, collision cell, modifiers. 7. Re-verify with fresh CCV and independent CRM.

7. Method Settings That Recover Signal Fast

  • ICP-MS. Increase RF power. Reduce nebulizer gas to stabilize aerosol. Enable kinetic energy discrimination. Choose isotopes free of overlaps.
  • ICP-OES. Select robust axial or radial view based on range. Optimize viewing height and integration time.
  • AAS. In flame mode, optimize fuel-oxidant ratio for maximum sensitivity without soot. In furnace mode, re-optimize pyrolysis and atomization with matrix modifiers.
  • CHNS. Increase oxygen pulse and combustion time. Replace reduction tube reagents. Verify drying and water trap performance to avoid bias on H.

8. Documentation That Prevents Repeat Failures

  • Embed recovery equations and acceptance limits in your LIMS templates.
  • Record instrument setpoints, gas flows, and maintenance actions with each batch.
  • Capture photo or lot codes for reagents and cones or reactors to link bias to component age.
  • Trend CCV and spike recoveries with control charts to detect slow drift before failure.

9. Worked Examples

# Metals spike example Unspiked sample = 0.80 mg/L Cu. Add 0.50 mg/L Cu spike. Measured spiked sample = 1.18 mg/L Cu. Spike Recovery = 100 × (1.18 − 0.80) / 0.50 = 76%. Diagnosis: Matrix suppression. Action: 5× dilution and matrix matching.
CHNS standard recovery example
Certified acetanilide: C = 71.09%.
Measured C = 67.5%.
Standard Recovery = 100 × 67.5 / 71.09 = 94.9% (acceptable).
If N is 85% of target simultaneously, prioritize reduction reactor condition or leaks.

10. Materials and Maintenance Checklist

ItemSpec or ActionFrequency
Type I water and high-purity acids.18 MΩ·cm, trace-metal grade acids.Each batch of standards.
Peristaltic pump tubing.Replace if flattened or >1 month old in routine use.Monthly or per hours of use.
Nebulizer and spray chamber.Acid soak and rinse. Verify uptake rate.Weekly or after high-TDS runs.
ICP-MS cones and torch.Inspect orifices. Remove salt and carbon deposits.Weekly or per manufacturer.
CHNS reactors and traps.Replace oxidation and reduction reagents. Refresh water traps.Per throughput or signal drift.
CRMs and check standards.Store cold and dark. Track expiry and lot.Each use.
Caution: Follow your chemical hygiene plan for oxidizers, HF, and pressurized gases. Use fume hoods, appropriate gloves, face shields, and vessel-compatibility charts.

FAQ

What recovery range should I accept if the method does not specify it?

Adopt 90–110% for standards and 80–120% for matrix spikes as default working limits. Tighten limits once your method shows stable performance.

How do I separate prep losses from instrument suppression?

Use a post-digestion or post-combustion spike. If recovery is low after prep but normal after the introduction system, the instrument is the dominant source of bias.

When should I use the method of standard additions?

Use it when matrix matching and internal standards do not correct suppression. It linearizes response in the actual matrix and is robust to signal depression.

Why does dilution improve recovery in ICP methods?

Dilution lowers total dissolved solids, viscosity, and acid strength. This improves aerosol generation and reduces ionization and transport suppression.

What is a rapid fix for CHNS incomplete combustion?

Reduce sample mass, add WO3 or V2O5, use tin capsules, and increase oxygen pulse. Verify reactor packing and leak integrity.