- Get link
- X
- Other Apps
- Get link
- X
- Other Apps
The purpose of this article is to provide a stepwise, expert method to diagnose and fix NMR shimming failures so you can recover narrow lines and stable baselines quickly.
1. Define success before you start
Set measurable targets for your probe and nucleus. Use a standard sample. Record line shape and lock stability.
| Metric | Typical Target (400 MHz, 5 mm, CDCl3) | How to Measure |
|---|---|---|
| FWHM of residual solvent peak | < 0.8 Hz | 1D single scan with long acquisition. |
| Line shape tails (±50 Hz) | < 2% of height | Line shape test with high digital resolution. |
| Lock drift over 5 min | < 2% | Monitor lock level vs time. |
| Auto-shim time | < 3 min | Standard vendor routine. |
2. Fix sample variables first
Shimming can only correct field inhomogeneity. It cannot fix poor samples. Control geometry, solvent, and temperature.
- Tube quality. Use premium 5 mm tubes with tight OD tolerance. Reject chipped or oval tubes.
- Fill height. Set 4.5–5.0 cm for 5 mm probes unless the probe manual states otherwise.
- Homogeneity. Filter out dust. Degas if bubbles form. Avoid suspended solids for routine 1D.
- Solvent choice. Use deuterated solvents with correct lock isotope. Avoid mixed deuteration unless validated.
- Salt and viscosity. High ionic strength or viscosity broadens lines. Dilute or change solvent.
- Temperature equilibrium. Equilibrate for 5–10 minutes in the magnet. Match spinner depth with a gauge.
Caution: Never shim on a sample that shows visible phase separation, precipitate, or bubbles. Replace the sample first.
3. Verify lock and basic setup
Lock failure often masquerades as shim failure. Confirm these steps in order.
- Center the sample at the calibrated depth. Level the spinner air. Check slow, stable spin if used.
- Tune and match the observe channel. Optimize lock gain and phase. Achieve a steady lock level.
- Zero-order shim reset. Load the instrument default shims for your probe and nucleus.
4. Run a fast shim triage
Use this quick decision tree to localize the fault.
| Symptom | Likely Cause | Action |
|---|---|---|
| Broad line with symmetric shape | Incorrect Z shims or fill height | Adjust Z1, Z2 only. Verify fill height and depth. |
| Asymmetric high-frequency tail | Transverse shims (X, Y) off | Run gradient XY shim or manual X/Y balancing. |
| Lock oscillates or drops | Temperature drift or bubbles | Equilibrate temperature. Degas or remake sample. |
| Auto-shim never converges | Wrong solvent, wrong nucleus, wrong profile | Load correct lock solvent and nucleus profile. Retry. |
| Great shim on standard, poor on sample | Sample heterogeneity or paramagnetics | Filter, dilute, or change solvent. Avoid steel needles and stir bars. |
5. Manual Z-only recovery
Recover from severe mis-shim by correcting Z terms first. Work from coarse to fine.
- Set X, Y, and higher orders to defaults. Do not touch until Z is acceptable.
- Adjust Z1 to maximize lock or minimize FWHM. Move slowly through the peak.
- Adjust Z2 to sharpen the top. Iterate Z1 and Z2 in small steps.
- If available, fine tune Z3 for tails. Stop when further change increases FWHM.
6. Gradient shimming workflow
Use pulsed field gradients to map and correct inhomogeneity. Select the correct routine for the probe and solvent.
- Verify gradient calibration and linearity on the service menu.
- Choose the 3D map routine for severe cases. Choose XY-only for drift correction.
- Limit maximum shim step size on sensitive cryoprobes.
- Run the routine twice if the first pass improves but does not converge.
Caution: Excessive gradient duty can warm the probe and degrade lock. Respect duty cycle limits for your hardware.
7. Baseline, phasing, and digital resolution
Shimming and processing interact. Verify acquisition and processing choices before blaming hardware.
- Acquisition time. Use long acquisition to resolve sub-Hz peaks.
- Digital resolution. Collect ≥ 64k points for line shape tests on 400–600 MHz.
- Window functions. Avoid heavy apodization during diagnostics.
- Zero filling and phase. Apply minimal zero filling and correct zero/first order phase.
8. Cryoprobe and HR-MAS specifics
Cryogenic probes demand smaller step sizes and longer equilibration. HR-MAS adds spinning sidebands that mimic tails.
- Reduce auto-shim aggressiveness. Increase averaging in lock optimization.
- Allow 10–15 minutes after temperature change. Watch lock stabilization.
- For HR-MAS, stabilize spin rate. Shim at the operating spin speed.
9. Environmental and instrument checks
Room and magnet conditions can dominate shim quality. Verify these before deep tuning.
- Temperature stability within ±1 °C. Avoid vents blowing on the magnet.
- Vibration sources removed. No carts or pumps touching the console or floor nearby.
- Magnet stray fields clear. Keep steel objects and tools away from the bore.
- Recent quenches or dumps recorded. If yes, perform full system homogeneity check.
10. Persistent failure checklist
Use this sequence when routines fail to converge.
# NMR shim recovery SOP 1. Load vendor default shim set for the probe and nucleus. 2. Insert a fresh reference sample (e.g., CHCl3 in CDCl3 or H2O in D2O). 3. Set fill height to 5.0 cm and spinner depth to the gauge value. 4. Optimize lock gain and phase. Confirm stable lock > 80%. 5. Tune and match the observe channel to within spec. 6. Run Z-only manual shim: iterate Z1 - Z2 - Z1. 7. Run gradient-based shim with conservative step sizes. 8. Verify line shape with long acquisition and high resolution. 9. Save the shim set with clear metadata (probe, temp, solvent). 10. Reinsert the user sample. Apply the saved shim set. Touch up XY only. 11. Common root causes and targeted fixes
| Root Cause | Diagnostic | Targeted Fix |
|---|---|---|
| Wrong shim file loaded | Metadata mismatch | Reload correct shim set for probe and nucleus. Lock again. |
| Probe detuned or mismatched | High reflected power | Retune and rematch. Check cables and connectors. |
| Gradient miscalibration | Map routine diverges | Run gradient calibration. Limit gradient amplitude. |
| Sample paramagnetics | Severe broadening on all peaks | Remove source. Use chelators only if compatible. Replace glassware. |
| Air bubbles | Lock jumps when spinning starts | Tap tube to release bubbles. Degas or remake sample. |
| Temperature mismatch | Lock drifts with time | Equilibrate longer. Verify VT airflow and sensor. |
| Spinner depth error | Shim good in one slot, not another | Re-gauge depth. Standardize the spinner turbine. |
12. Save and version-control good shims
Store successful shim sets with metadata. Reuse as a starting point.
- Include probe ID, nucleus, solvent, temperature, and sample height.
- Record date and operator. Add notes on gradients and routines used.
- Archive separate sets for unusual solvents or viscous matrices.
FAQ
Should I shim on the analyte peak or the lock?
Shim against the lock signal for general work. For demanding experiments, confirm line shape on the target peak after lock-based shimming.
Is manual shimming still useful with modern auto routines?
Yes. Manual Z1 and Z2 adjustments often recover severe cases faster. Use auto routines to refine after Z terms are close.
Why do shims fail after changing temperature?
Field and susceptibility change with temperature. Allow time for equilibrium and repeat a brief shim touch-up.
Can spinning improve shimming?
Slow spinning can average minor inhomogeneity but can also destabilize lock if bubbles exist. Use only after the static shim is acceptable.
What line width should I accept?
Follow your instrument specification and matrix. Aim for sub-Hz FWHM on standards. Heavier solvents or salts may not reach that value.
- Get link
- X
- Other Apps