- Get link
- X
- Other Apps
- Get link
- X
- Other Apps
The purpose of this article is to deliver proven methods to improve pipette accuracy in laboratories using best practices, quick calibration checks, and robust quality control workflows.
1. Understand what drives pipette inaccuracy.
Accuracy errors arise from equipment limits, operator technique, environment, and consumables. Treat each as a control point.
| Error Source | Mechanism | High-leverage Control |
|---|---|---|
| Temperature | Air cushion expansion shifts delivered volume. | Equilibrate pipette, tips, and liquid to 20–25°C for 30 minutes. |
| Tip fit | Leakage or dead volume from poor seal. | Use manufacturer-matched tips and firmly seat with a straight push. |
| Aspiration speed | Turbulence and incomplete filling. | Use smooth, slow aspiration and a 1–2 second post-aspiration hold. |
| Immersion depth | Too deep increases hydrostatic head. | Immerse 2–3 mm for ≤200 µL and 3–6 mm for >200 µL. |
| Angle | Off-axis changes meniscus shape. | Keep the pipette vertical during aspiration and ≤20° tilt during dispense. |
| Evaporation | Mass loss during weighing or pre-wetting gaps. | Pre-wet tips 3× and minimize balance exposure time. |
| Wear and seals | Piston seal friction and leaks. | Replace seals and lubricate per OEM schedule. |
Caution: Never wipe the outside of a filled tip. Capillary action will remove liquid and bias the volume low.
2. Match the tool to the task.
Use the smallest pipette that spans the target volume range. Stay in the 35–100% range of the pipette’s nominal capacity for best accuracy. Use low-retention tips for viscous or low-surface-tension liquids. Use positive-displacement tips for volatile, viscous, or foaming samples.
Caution: For dense or volatile liquids, calibration must be verified with the actual matrix or a validated correction factor.
3. Standardize a high-accuracy technique.
Apply a consistent motion to remove user-to-user variability.
| Step | Best Practice | Why it Works |
|---|---|---|
| Pre-wetting | Pre-wet the tip 3× with the sample. | Conditions the air cushion and reduces evaporation error. |
| Immersion | 2–3 mm depth, vertical shaft. | Minimizes hydrostatic head and sidewall effects. |
| Aspiration | Slow to first stop, hold 1–2 s. | Allows meniscus stabilization and full tip filling. |
| Withdrawal | Withdraw along the vessel wall. | Prevents droplet loss from tip exterior. |
| Dispense | Touch off on wall, then second stop. | Removes residual film and ensures full delivery. |
| Post-dispense | Maintain contact 1 s before removal. | Prevents backflow and droplet formation. |
4. Control the environment and ergonomics.
Work near 20–25°C and 45–60% relative humidity. Avoid drafts from HVAC or people traffic near the balance. Keep samples covered between steps. Reduce hand heat transfer by using brief handling or a pipette stand. Rotate tasks to limit fatigue and grip variation.
5. Verify performance with a quick gravimetric check.
Use a calibrated analytical balance with 0.01 mg or better for ≤200 µL and 0.1 mg for higher volumes. Use a weighing vessel with evaporation trap or lid. Tare with vessel and water at test temperature.
# Gravimetric verification (water at 20–25°C) Given: m_i = mass after dispense i (g) m_0 = mass before any dispense (g) ρ_w = density of water at test temperature (g/mL) Z = air buoyancy correction factor (unitless, typically ~1.0012 near 20–25°C) E = evaporation correction per reading (mL), if applicable
Per-dispense delivered volume:
V_i = ((m_i − m_{i−1}) * Z) / ρ_w − E
Systematic error (%E):
%E = 100 * (mean(V_i) − V_set) / V_set
Repeatability (CV):
CV% = 100 * (sd(V_i) / mean(V_i))
Acceptance:
Define internal limits based on ISO 8655 class for the pipette size, or tighter if your method requires it.
Caution: Record ambient temperature for each run. Use the matching water density value for that temperature to avoid bias.
6. Schedule calibration and preventive maintenance.
Set monthly user checks at low, mid, and high volumes. Set quarterly or semiannual full calibrations with at least 10 replicates per test point. Replace seals and o-rings on a fixed schedule or when plunger force increases. Label pipettes with the last calibration date, due date, and current status.
7. Handle special liquids correctly.
| Liquid Type | Adjustment | Tip/Pipette Choice |
|---|---|---|
| Viscous (glycerol, syrups) | Slower aspiration and dispense, extend hold 2–3 s. | Low-retention or positive-displacement. |
| Volatile (acetone, hexane) | Pre-wet 5×, minimize delay, verify gravimetrically. | Positive-displacement preferred. |
| Foaming (detergents) | Aspirate very slowly, dispense to vessel wall. | Air-cushion with low-retention tips. |
| Hot or cold | Equilibrate, apply matrix-specific correction factor. | Positive-displacement if large bias persists. |
8. Build a simple daily QC chart.
Track mean, %E, and CV% at a fixed setpoint. Use warning limits at 2 SD and action limits at 3 SD from your historical mean. Investigate any trend across days or operators.
9. Quick SOP for high-precision pipetting.
1. Select the smallest pipette that covers the target volume. 2. Install matched tips; push straight without rocking. 3. Pre-wet the tip 3 times with the sample. 4. Immerse 2–3 mm, hold vertical, aspirate slowly to first stop. 5. Hold 1–2 s; withdraw along the wall. 6. Dispense to vessel wall; hold 1 s; press to second stop. 7. Touch off; retract; discard tip without wiping. 8. Log lot numbers for tips and the pipette ID used. 9. Run a 5-rep gravimetric check at session start for critical assays. Quote
"Measure what is measurable, and make measurable what is not.".
FAQ
How many pre-wet cycles are enough?
Three cycles stabilize the air cushion for most aqueous samples. Use five for volatile or highly wetting liquids.
Should I use reverse pipetting?
Use reverse pipetting for viscous or foaming samples to improve precision and reduce bubbles.
How often should I recalibrate after a failure?
After any out-of-tolerance result, service the pipette and perform a full calibration before returning it to use.
What balance class do I need?
Use a 0.01 mg balance for microvolumes up to 200 µL and a 0.1 mg balance for larger volumes.
- Get link
- X
- Other Apps